| Master-8 is an eight channel general purpose stimulator for nerve and muscle stimulation procedures: |
All eight channels can be operated independently, or synchronized to produce complex patterns of pulses.
Use some channels to stimulate your preparation, while using other channels to trigger other instruments (oscilloscope, PC, etc.) in synchronization with the stimulation.
| Master-8 is an extremely flexible system since: |
As a stand-alone unit – All parameters are set by the front panel keys. You can switch between 8 stored experiments; each using all the eight channels.
When connected to a PC (via USB) – Download/upload new experiments. Your PC controls and modifies the parameters of Master-8 in accordance with the feedback from the experiment.
You can use the powerful scheduler in order to modify parameters and switch between experiments at scheduled times.
Download our new Master-8 SDK
(Software Development Kit) – it is free!
|
Now you can control the Master-8 in 3 ways:
| 1. |
As a stand alone unit - via its front panel keys.
|
| 2. |
Via your PC – using our Master-8 Control Software
|
| 3. |
From within your own application/program (MATLAB, LABVIEW etc.) –
using our new Master-8 SDK. |
More about the Master-8 SDK
Download the Master-8 SDK
| Master-8 is an extremely reliable system as: |
Wide time range – interval time between pulses/trains from 60 msec. (16,666 PPS) to 3999 sec. (1/3999 PPS).
4 digits resolution – For example: Set the pulse delay to be 10.01 millisec.
You can modify parameters while stimulating – you don’t have to stop the experiment.
For example: while stimulating, switch the stimulation rate from one range to another (e.g. from 98 PPS to 102 PPS).
Simple switch-off the stimulation – important for cases where immediate stimulation cut off is required.
Each channel can operate in any of the following modes:
Free-Run: The channel delivers repetitive pulses.
Trig: Single pulse after triggering (manually, internally from other channels, or externally).
Train: Train of pulses after triggering.
DC: The channel delivers a DC output and is time independent, it can be turned on or off manually
Gate: Each time that an external signal is applied, the channel delivers repetitive pulses. Whenever the external input stops, the channel stops working.
| Inter-Channel Connections |
Each channel can be internally connected to any other channel with no need to connect wires between channels. Each channel can be internally connected to many channels simultaneously.
Each channel operates within the following ranges:
|
|
MIN
|
MAX
|
|
|
Number of Pulses per Train
|
1
|
59,990
|
|
|
Pulse Duration
|
40 microsec.
|
3,999 sec
|
8 decades |
|
Delay Time
|
100microsec.
|
3,999 sec
|
> 7 decades |
|
Interval Time (1/rate)
|
60 microsec.
|
3,999 sec
|
> 7 decades |
Each channel can deliver output pulses with amplitudes ranging from -10V to +10V and with maximum current of 20mA.
Three additional outputs deliver combined multilevel pulses:
|
Out 2+3
|
Bi-level pulses (e.g. +/- pulses).
|
|
Out 6+7+8
|
Tri-level pulses.
|
|
Combined Output
|
up to 7 level pulses.
|
In addition to its standard 8 channels, Master-8 features 2 internal clocks:
The first clock counts the time in seconds up to 24 hours and is used to measure the time elapsed from an event (e.g. the beginning of the experiment). The second clock counts the time in tenths of seconds up to 1 hour and is used to measure intermediate times.
Master-8 can be mounted in a standard 19-inch rack.
Regulatory Compliance: UL and CE
Master-8 has not been approved for direct patient use.